Appendix H: Evidence tables


 Bibliographic details   Study type &   Patient characteristics   Test, Reference Standard,   Results   Reviewers Comments
 Evidence level   Threshold for a positive test
 France, Italy,   GA: > 36 week = 80.2%      Sternum
 Switzerland)   Postnatal age: < 48 hours = 16.3%   Blood sample also collected for   r = 0.85, P < 0.001
    Gender: Not reported   TSB estimation using HPLC-B
 77       technique at the same time   Correlation of lab TSB levels with TSB
    Ethnicity      levels using HPLC-B
    White = 66.7%   Diagnostic accuracy of TcB   (Pearson correlation coefficient, n = 210)
 Asian = 14.8%   (various thresholds) estimated at
 Hispanic = 6.7%   various thresholds and plotted on   r = 0.93, P < 0.001
 Other = 11.9%   ROC curve.
               Bland Altman analysis for difference
 Exclusion: not defined    between lab TSB and TcB

               Forehead
               MD = +2.4 micromol/litre (95% CI -2.4 to
               +7.1)
               SD = 35.4 micromol/litre

               Sternum
               MD = -14.8 micromol/litre (95% CI -19.9 to
               +9.5)
               SD = 38.4 micromol/litre

               Diagnostic accuracy of TcB on forehead
               (threshold 187 micromol/litre) for detecting
               TSB > 222 micromol/litre by HLPC-B

               Sensitivity: 93%
               Specificity: 73%

               Diagnostic accuracy of TcB (threshold
               238 micromol/litre) for detecting TSB >
               290 micromol/litre by HLPC-B

               Sensitivity: 90%
               Specificity: 87%


 Boo NY;   Study Type:   Healthy term Malaysian babies with   TcB using BiliChek from the   Correlation of TcB levels with lab TSB   Unselected population
    Diagnostic study   hyperbilirubinaemia   forehead and midpoint of sternum –   levels (Pearson correlation coefficient,   Test & Reference test described
 Year: 2007         number of measurements from each   n = 345)   adequately
    Evidence Level:   n = 345   site not specified      Test and reference test carried
 Country:   Ib   mean BW: 3056 ± 487 g,      Forehead   out within one hour
 Malaysia      median GA 38 weeks   Reference standard: Laboratory   All babies   Blinding – yes
       TSB levels within 30 minutes of   r = 0.80, P < 0.0001   Data not given for the mean
 78    postnatal age: range 9 – 388   TcB measurement   Malays: r = 0.79, P < 0.0001   difference and SD from Bland
    Gender: Males = 60%      Chinese: r = 0.84, P < 0.0001   Altman analysis for TSB – TcB

    Ethnicity   Diagnostic accuracy of TcB   Indians: r = 0.83, P < 0.0001
 Malays = 63.8%   (various thresholds) calculated for