Page 330 - 16Neonatal Jaundice_compressed
P. 330

Neonatal jaundice


              Bibliographic details   Study type &   Patient characteristics   Test, Reference Standard,   Results                Reviewers Comments
                                Evidence level                            Threshold for a positive test
                                             Chinese = 30.7%              detecting TSB > 250, > 280, and   Sternum
                                             Indians = 5.5%,              > 300 micromol/litre     All babies
                                                                                                   r = 0.86, P < 0.0001
                                                                                                   Malays: r = 0.86, P < 0.0001
                                             Prevalence of TSB >                                   Chinese: r = 0.86, P < 0.0001
                                             300 micromol/litre = 27.5% (95/345)                   Indians: r = 0.94, P < 0.0001

                                             Exclusion: infants who had received                   Correlation of TcB levels with lab TSB
                                             phototherapy or exchange transfusion,                 levels depending on the time of
                                             congenital anomalies,                                 measurement
                                             severely ill,                                         (Pearson correlation coefficient, 79% of
                                             foreigners,                                           infants with TSB > 300 had measurement at
                                             those with conjugated                                 > 80 hours)
                                             hyperbilirubinaemia.
                                                                                                   At < 80 hours
                                                                                                   r = 0.85, P < 0.001

                                                                                                   At > 80 hours
                                                                                                   r = 0.71, P < 0.001

                                                                                                   Diagnostic accuracy of TcB for detecting
                                                                                                   TSB > 300 micromol/litre

                                                                                                   Forehead (threshold 250 micromol/litre)
                                                                                                   Sensitivity: 100%
                                                                                                   Specificity: 39.2%

                                                                                                   Forehead (threshold 260 micromol/litre)
                                                                                                   Sensitivity: 75.8%
                                                                                                   Specificity: 84.8%

                                                                                                   Sternum (threshold 200 micromol/litre)
                                                                                                   Sensitivity: 100%
                                                                                                   Specificity: 33.6%

                                                                                                   Sternum (threshold 280 micromol/litre)
                                                                                                   Sensitivity: 92.6%
                                                                                                   Specificity: 84%


              Ebbesen F;        Study Type:   All newborns more than 24 hours old   TcB measurement using BiliChek   Correlation of TcB levels with lab TSB   Unselected population
                                Diagnostic study   who for clinical reasons had their plasma   from forehead, sternum, knee and   levels (Pearson correlation coefficient,   Test & Reference test described
              Year: 2002                     bilirubin determination during the day,   the foot – mean of 5 measurements   n = 210)   adequately
                                Evidence Level:   except at weekends.     from each site taken for data                           Test and reference test carried
              Country:          III                                       analysis.                Group 1:                       out within one hour
              Denmark                        Group 1: Both preterm infants                         Forehead                       Blinding – not specified
                                             < 35 weeks and sick term and near-term   Reference standard: Laboratory   r = 0.88, p > 0.05   Data not given for the mean
              45                             infants in the NICU          TSB levels taken concurrently with   Sternum            difference and SD from Bland



            238
   325   326   327   328   329   330   331   332   333   334   335